In this study, a liquid chromatographic method was developed for the fast determination of lincomycin, polymyxin and vancomycin in a preservation solution for transplants. A Kinetex EVO C18 (150 × 4.6 mm, 2.6 μm) column was utilized at 45 ◦C. Gradient elution was applied using a mixture of mobile phases A and B, both including 30 mM phosphate buffer at pH 2.0 and acetonitrile, at a ratio of 95:5 (v/v) for A and 50:50 (v/v) for B. A flow rate of 1.0 mL/min, an injection volume of 20 μL and UV detection at 210 nm were used. A degradation study treating the three antibiotics with 0.5 M hydrochloric acid, 0.5 M sodium hydroxide and 3% H2O2 indicated that the developed method was selective toward lincomycin, polymyxin, vancomycin and their degradation products. Other ingredients of the preservation solution, like those from the cell culture medium, did not interfere. The method was validated with good sensitivity, linearity, precision and accuracy. Furthermore, lincomycin, polymyxin and vancomycin were found to be stable in this preservation solution for 4 weeks when stored at −20 ◦C.